14.06.2014 Views

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Search Functions - Analyzer<br />

ESIB<br />

The following example shows the effect of different PEAK EXCURSION<br />

settings.<br />

0<br />

-10<br />

-20<br />

Marker [T1]<br />

199.0140 MHz<br />

-22.4 dBm<br />

1<br />

1<br />

2<br />

4<br />

-30<br />

3<br />

-40<br />

-50<br />

42dB<br />

30dB<br />

46dB<br />

-60<br />

-70<br />

-80<br />

-90<br />

-100<br />

Fig. 4-12 Example for level measurements at various peak excursion settings<br />

Maximum relative level change of the measured signals:<br />

Signal 2: 42dB<br />

Signal 3 30dB<br />

Signal 4: 46dB<br />

With Peak Excursion 40dB signals 2 and 4 are detected by NEXT PEAK or<br />

NEXT PEAK RIGHT. Signal 3 is not detected since it is only decreased by 30<br />

dB before rising again.<br />

Order in which signals are found:<br />

PEAK: Signal 1 or PEAK: Signal 1<br />

NEXT PEAK: Signal 2 NEXT PEAK RIGHT: Signal 2<br />

NEXT PEAK: Signal 4 NEXT PEAK RIGHT: Signal 4<br />

With Peak Excursion 20dB signal 3 is detected as well since its highest level<br />

change of 30 dB is now higher than the peak excursion.<br />

Order in which signals are found:<br />

PEAK: Signal 1 or PEAK: Signal 1<br />

NEXT PEAK: Signal 2 NEXT PEAK RIGHT: Signal 2<br />

NEXT PEAK: Signal 4 NEXT PEAK RIGHT: Signal 3<br />

NEXT PEAK: Signal 3 NEXT PEAK RIGHT: Signal 4<br />

1088.7531.12 4.206 E-15

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!