ESIB<br />

Switchover to Remote Control<br />

Remote Control via IEC Bus<br />

Setting the Device Address<br />

In order to operate the instrument via the IEC-bus, it must be addressed using the set IEC-bus address.<br />

The IEC-bus address of the instrument is factory-set to 20. It can be changed manually in the SETUP -<br />

GENERAL SETUP menu or via IEC bus. Addresses 0 to 31 are permissible.<br />

<strong>Manual</strong>ly:<br />

½ Call SETUP - GENERAL SETUP menu<br />

½ Enter desired address in table GPIB ADDRESS<br />

½ Terminate input using one of the unit keys (=ENTER).<br />

Via IEC bus:<br />

CALL IBFIND("DEV1", receiver%)<br />

’Open port to the instrument<br />

CALL IBPAD(receiver%, 20)<br />

’Inform controller about old address<br />

CALL IBWRT(receiver%, "SYST:COMM:GPIB:ADDR 18")’Set instrument to new address<br />

CALL IBPAD(receiver%, 18)<br />

’Inform controller about new address<br />

Return to <strong>Manual</strong> Operation<br />

Return to manual operation is possible via the front panel or the IEC bus.<br />

<strong>Manual</strong>ly:<br />

½ Press the LOCAL key.<br />

Notes: – Before switchover, command processing must be completed as<br />

otherwise switchover to remote control is effected immediately.<br />

– The LOCAL key can be disabled by the universal command LLO<br />

(see chapter 8) in order to prevent unintentional switchover. In<br />

this case, switchover to manual mode is only possible via the IEC<br />

bus.<br />

– The LOCAL key can be enabled again by deactivating the REN<br />

line of the IEC bus (see chapter 8).<br />

Via IEC bus: ...<br />

CALL IBLOC(receiver%)<br />

...<br />

’Set instrument to manual operation.<br />