ESIB<br />

SENSe Subsystem<br />

:[SENSe:]ADEMod:SQUelch:LEVel 30 to -150 dBm<br />

This command defines the switching threshold for the squelch referred to the measured signal.<br />

Example: ":ADEM:SQU -10DBM"<br />

Features: *RST value: -40dBm<br />

SCPI: device-specific<br />

Mode:<br />

VA-A<br />

:[SENSe:]ADEMod:SBANd NORMal | INVerse<br />

This command selects the side band for the demodulation.<br />

Example: ":ADEM:SBAN INV"<br />

Features: *RST value: NORMal<br />

SCPI: device-specific<br />

Mode:<br />

VA-A<br />

:[SENSe:]ADEMod:RTIMe ON | OFF<br />

This command selects whether the demodulation performed in real time or in blocks.<br />

Example: ":ADEM:RTIM ON"<br />

Features: *RST value: ON<br />

SCPI: device-specific<br />

Mode:<br />

VA-A<br />