ESIB<br />

Analyzer - Frequency and Span<br />

Stop Frequency – STOP Key<br />

FREQUENCY – STOP menu:<br />

FREQUENCY<br />

CENTER/<br />

FREQ<br />

SPAN/<br />

ZOOM<br />

START STOP<br />

STOP FREQ<br />

STOP<br />

MANUAL<br />

START<br />

FIXED<br />

CENTER<br />

FIXED<br />

SPAN<br />

FIXED<br />

The STOP key opens a menu which displays the various<br />

options for setting the stop frequency of the sweep.<br />

The STOP MANUAL softkey is automatically active and opens<br />

the entry window for manual input of the stop frequency. At the<br />

same time, parameter coupling is set to START FIXED.<br />

The START FIXED, CENTER FIXED and SPAN FIXED<br />

softkeys are mutually exclusive selection switches. Only one<br />

switch can be active at any one time. The softkeys are used to<br />

select the frequency coupling. The frequency coupling defines<br />

which of the dependent parameters start frequency, center<br />

frequency or span is to remain constant when the stop<br />

frequency is changed.<br />

FREQ AXIS<br />

LIN LOG<br />

STOP<br />

MANUAL<br />

The STOP MANUAL softkey activates the manual entry window for the stop<br />

frequency.<br />

The allowed range of values for the stop frequency is:<br />

minspan ≤ f stop ≤ f max f stop stop frequency<br />

minspan smallest selectable span (10Hz)<br />

f max max. frequency<br />

IEC/IEEE-bus command :[SENSe:]FREQuency:STOP 13 GHz<br />

START<br />

FIXED<br />

If the START FIXED softkey is active, the start frequency remains constant<br />

when the stop frequency is changed. The center frequency is adjusted to the<br />

new frequency range. The START FIXED coupling is the default setting.<br />

IEC/IEEE-bus command :[SENSe:]FREQ:STOP:LINK STARt<br />

CENTER<br />

FIXED<br />

If the CENTER FIXED softkey is active, the center frequency remains<br />

constant when the stop frequency is changed. The start frequency is adjusted<br />

to the new frequency range.<br />

IEC/IEEE-bus command :[SENSe:]FREQ:STOP:LINK CENTer<br />

SPAN<br />

FIXED<br />

If the SPAN FIXED softkey is active, the span remains constant when the<br />

stop frequency is changed. The start frequency is adjusted to the new<br />

frequency range.<br />

IEC/IEEE-bus command :[SENSe:]FREQ:STOP:LINK SPAN<br />

FREQ AXIS<br />

LIN LOG<br />

The FREQ AXIS LIN/LOG softkey switches between linear and logarithmic<br />

scales on the frequency axis (see START key).<br />

IEC/IEEE-bus command :[SENSe:]SWEep:SPACing LIN | LOG<br />

1088.7531.12 4.157 E-15