14.06.2014 Views

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Main Markers - Analyzer<br />

ESIB<br />

If a marker (or, delta marker) necessary for a marker function is not available, it will be automatically<br />

checked whether or not the enabling of the corresponding marker is possible (see above). If this is not<br />

the case, a warning is issued.<br />

WARNING:<br />

No trace active<br />

The activation of the desired marker function is then not possible.<br />

On the other hand, if the marker can be enabled, a peak search is automatically performed. Thereafter,<br />

the desired marker function can be executed.<br />

MARKER NORMAL menu:<br />

ALL MARKER<br />

OFF<br />

The ALL MARKER OFF softkey switches off all markers (reference and delta<br />

markers). Similarly, it switches off all functions and displays correlated with<br />

the markers/delta markers (signal count, signal track, marker zoom, N dB<br />

down, shape factor, marker list and marker info).<br />

IEC/IEEE-bus command :CALCulate:MARKer:AOFF<br />

SIGNAL<br />

TRACK<br />

After each sweep, the SIGNAL TRACK softkey starts a search for the<br />

maximum signal level on the screen (PEAK SEARCH) and then sets the<br />

center frequency to the frequency of this signal (MARKER ->CENTER).<br />

If a threshold line is enabled, only signals above the threshold level are<br />

considered. If no signal is over the threshold, the center frequency remains<br />

constant until a signal is available which is over the threshold.<br />

If no marker is switched to the active trace, the next unused marker is<br />

automatically activated and positioned by the peak search routine.<br />

In order to reduce the search range, it is possible to apply the SEARCH LIMIT<br />

ON/OFF function (see section: "Search Functions"). If the center frequency is<br />

changed, the position of the frequency line relative to the center frequency<br />

remains the same, ie its absolute position is correspondingly changed.<br />

The softkey is only available in the frequency domain (spectrum) display<br />

mode (span > 0.)<br />

IEC/IEEE-bus commands<br />

:CALCulate:MARKer:COUNt ON |OFF;<br />

:CALCulate:MARKer:COUNt:FREQuency?<br />

1088.7531.12 4.178 E-15

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!