Configuration/Setup<br />

ESIB<br />

Parity – Bit parity check<br />

NONE no parity check (default setting)<br />

EVEN even parity check<br />

ODD odd parity check<br />

PARITY<br />

NONE<br />

EVEN<br />

ODD<br />

IEC/IEEE-bus command<br />

:SYSTem:COMM:SERial[:RECeive]:PARity[:TYPE]<br />

EVEN| ODD| NONE<br />

Stop bits – Number of stop bits<br />

Available are 1 and 2. The default setting is 1 stop bit.<br />

STOPBITS<br />

1<br />

2<br />

IEC/IEEE-bus command<br />

:SYSTem:COMMunicate:SERial[:RECeive]:SBITs 1|2<br />

HW-Handshake – Hardware handshake protocol<br />

The integrity of data transmission can be improved through the use of a<br />

hardware handshake protocol which prevents uncontrolled transmission of<br />

data and the resulting possible loss of data bytes. With the hardware<br />

handshaking procedure, additional interface lines are used to transmit<br />

acknowledge signals with which the data transmission can be controlled and,<br />

if necessary, stopped until the receiver is ready to receive data again.<br />

A prerequisite for handshaking is, however, that the interface lines (DTR and<br />

RTS) are provided between the sender and receiver. For a simple, 3-wire<br />

connection, this is not the case and a hardware handshake cannot be<br />

realized.<br />

Default setting is NONE.<br />

HW-HANDSHAKE<br />

None<br />

DTR/RTS<br />

IEC/IEEE-bus command<br />

:SYSTem:COMM:SERial:CONTrol:DTR IBFull|OFF<br />

:SYSTem:COMM:SERial:CONTrol:RTS IBFull|OFF<br />