SENSe Subsystem<br />

ESIB<br />

SENSe:BANDwidth Subsystem<br />

This subsystem controls the setting of the instrument’s filter bandwidths. Both groups of commands<br />

(BANDwidth and BWIDth) perform the same functions.<br />

COMMAND PARAMETERS UNIT COMMENT<br />

[SENSe]<br />

:BANDwidth | :BWIDth<br />

[:RESolution]<br />

:AUTO<br />

:MODE<br />

:FFT<br />

<br />

<br />

ANALog|DIGital<br />

<br />

HZ<br />

--<br />

--<br />

--<br />

:RATio<br />

:FILTer<br />

:AUTO<br />

:RATio<br />

:DEMod<br />

:COUPling<br />

:PLL<br />

:VIDeo<br />

<br />

<br />

<br />

| SINe | PULSe |<br />

NOISe<br />

<br />

<br />

AUTO | HIGH | MEDium | LOW<br />

--<br />

DBHZ<br />

--<br />

--<br />

HZ<br />

Vector Signal Analysis<br />

Vector Signal Analysis<br />

:[SENSe:]BANDwidth|BWIDth[:RESolution] 1 Hz to 10 MHz<br />

This command defines the IF bandwidth (6-dB bandwidth) of the receiver or resolution bandwidth of<br />

the analyzer.<br />

Example:<br />

":BAND 1MHz"<br />

Features: *RST value: - (AUTO is set to ON)<br />

SCPI: conforming<br />

Modes:<br />

R, A, VA<br />

In the receiver mode, ESIB offers the IF bandwidths (6 dB bandwidths) 10 Hz, 100 Hz, 200 Hz, 1<br />

kHz, 9 kHz, 10 kHz, 100 kHz, 120 kHz, 1 MHz and 10 MHz. The bandwidth is limited by the set<br />

receive frequency: RES BW ≤ f E /2. When the quasi peak detector is switched on, a fixed bandwidth<br />

is preset depending on the frequency (see also chapter 4, section "Selecting a Detector").<br />

In the analyzer and vector analyzer mode, the values for the resolution bandwidth are rounded in<br />

1 | 2 | 3 | 5 steps.<br />

:[SENSe:]BANDwidth|BWIDth[:RESolution]:AUTO ON | OFF<br />

In receiver mode with activated quasi peak detector, this command either automatically couples the<br />

IF bandwidth of the receiver to the frequency range or cancels the coupling .<br />

In analyzer mode, this command either automatically couples the resolution bandwidth of the<br />

instrument to the span or cancels the coupling.<br />

Example:<br />

":BAND:AUTO OFF"<br />

Features: *RST value: ON<br />

SCPI: conforming<br />

Modes:<br />

R, A, VA<br />

The automatic coupling matches the resolution bandwidth to the currently set span according to the<br />

relationship between span and resolution bandwidth.<br />

1088.7531.12 6.106 E-15