ESIB<br />

Configuration/Setup<br />

Enabling Firmware Options<br />

CONFIGURATION SETUP menu:<br />

OPTIONS<br />

The OPTIONS softkey opens a submenu for entering keywords for new<br />

firmware options (Application Firmware Modules). Options which are already<br />

available are indicated in a table that is opened when the menu is called.<br />

ENABLE NEW<br />

OPTION<br />

The ENABLE OPTION softkey activates the entry of a<br />

keyword for a firmware option.<br />

One or several keywords may be specified in the entry<br />

field. When a valid keyword is entered, the message<br />

OPTION KEY OK appears in the info line and the option<br />

is written into the FIRMWARE OPTIONS table.<br />

The FIRMWARE OPTIONS table can also be displayed<br />

using the FIRMWARE OPTIONS softkey in the INFO<br />

menu.<br />

If a keyword is invalid the message OPTION KEY<br />

INVALID is displayed in the info line.<br />

IEC/IEEE-bus command --<br />

External Reference Oscillator<br />

In general, the ESIB can generate all internal oscillators from either the internal or an external reference<br />

oscillator. A 10 MHz crystal oscillator is used as the internal reference source. A reference oscillator<br />

output is also available at the rear panel via the EXT REF IN/OUT connector.<br />

This output can be used, e.g., to synchronize other instruments to the ESIB. This connector can also be<br />

used as an input connector for application of an external frequency standard. The ESIB requires that the<br />

frequency of the external standard be entered so that the internal oscillators can be synchronized to it.<br />

CONFIGURATION SETUP Menu:<br />

REFERENCE<br />

INT EXT<br />

The REFERENCE INT EXT softkey toggles between the internal and external<br />

references.<br />

IEC/IEEE-bus command<br />

:[SENSe:]ROSC:SOURce INT | EXT<br />

EXT REF<br />

FREQUENCY<br />

The EXT REF FREQUENCY softkey activates the entry of the frequency of<br />

the external source.<br />

Range of input values is 1 MHz to 16 MHz in 1 MHz steps.<br />

IEC/IEEE-bus command :[SENSe:]ROSC:EXT:FREQ 13 MHz<br />

1088.7531.12 4.37 E-15