ESIB<br />

SENSe Subsystem<br />

:[SENSe:]CORRection:TSET[:STATe] ON | OFF<br />

This command switches the selected transducer set on or off.<br />

Example: ":CORR:TSET ON"<br />

Features: *RST value: OFF<br />

SCPI: device-specific<br />

Modes:<br />

R, A<br />

Prior to this command, the command SENS:CORR:TSET:SEL must be sent.<br />

:[SENSe:]CORRection:TSET:DELete<br />

This command deletes the selected transducer set.<br />

Example: ":CORR:TSET:DEL"<br />

Features: *RST value: -<br />

SCPI: device-specific<br />

Modes:<br />

R, A<br />

This command is an event and thus has no *RST value assigned.<br />

Prior to this command, the command SENS:CORR:TSET:SEL must be sent.<br />

:[SENSe:]CORRection:CVL:CATalog?<br />

This command polls the names of all conversion-loss tables stored on the harddisk.<br />

Syntax of output format:<br />

,,<br />

,,,,,,....,,<br />

<br />

Example: ":CORR:CVL:CAT?"<br />

Feature: *RST value: -<br />

SCPI: device-specific<br />

Mode:<br />

A<br />

This command is only valid in conjunction with option External Mixer Output, FSE-B21.<br />

:[SENSe:]CORRection:CVL:SELect <br />

This command selects the Conversion Loss Table designated with . If is<br />

not available, a new Conversion Loss Table will be created.<br />

Parameter: ::= Name of Conversion Loss Table as string data with<br />

a maximum of 8 characters<br />

Example: ":CORR:CVL:SEL ’LOSS_TAB’"<br />

Features: *RST value: -<br />

SCPI: device-specific<br />

Mode:<br />

A<br />

This command must be sent prior to the subsequent commands used to change/activate the<br />

Conversion Loss files. It is only valid in conjunction with option External Mixer Output, FSE-B21.<br />