ESIB<br />

Introduction<br />

5 Remote Control - Basics<br />

In this chapter you find:<br />

• instructions how to put the ESIB into operation via remote control,<br />

• a general introduction to remote control of programmable instruments. This includes the description<br />

of the command structure and syntax according to the SCPI standard, the description of command<br />

execution and of the status registers,<br />

• diagrams and tables describing the status registers used in the ESIB.<br />

In chapter 6, all remote control functions are described in detail. The subsystems are listed by<br />

alphabetical order according to SCPI. All commands and their parameters are listed by alphabetical<br />

order in the command list at the end of chapter 6.<br />

Program examples for the ESIB can be found in chapter 7.<br />

The remote control interfaces and their interface functions are described in chapter 8.<br />

Introduction<br />

The instrument is equipped with an IEC-bus interface according to standard IEC 625.1/IEEE 488.2 and<br />

two RS-232 interfaces. The connector is located at the rear of the instrument and permits to connect a<br />

controller for remote control.<br />

The internal controller function may also be used as a controller (see chapter 1, section "Controller<br />

function).<br />

In addition, the instrument is equipped with an RSIB interface that allows instrument control by Visual<br />

C++ and Visual Basic programs<br />

The instrument supports the SCPI version 1994.0 (Standard Commands for Programmable<br />

Instruments). The SCPI standard is based on standard IEEE 488.2 and aims at the standardization of<br />

device-specific commands, error handling and the status registers (see section "SCPI Introduction").<br />

This section assumes basic knowledge of IEC-bus programming and operation of the controller. A<br />

description of the interface commands is to be obtained from the relevant manuals. The RSIB interface<br />

functions are matched to the function interface for IEC/IEEE-bus programming from National<br />

Instruments. The functions supported by the DLLs are listed in chapter 8.<br />

The requirements of the SCPI standard placed on command syntax, error handling and configuration of<br />

the status registers are explained in detail in the respective sections. Tables provide a fast overview of<br />

the commands implemented in the instrument and the bit assignment in the status registers. The tables<br />

are supplemented by a comprehensive description of every command and the status registers. Detailed<br />

program examples of the main functions are to be found in chapter 7.<br />

The program examples for IEC-bus programming are all written in Quick BASIC.<br />