ESIB<br />

Analyzer - Main Markers<br />

CHANNEL<br />

SPACING<br />

The CHANNEL SPACING softkey opens an table for defining the channel<br />

spacing. The channel spacing is the spacing of the center frequency of a<br />

given channel relative to the center of the transmission channel.<br />

The default setting for channel spacing is 20 kHz.<br />

CHANNEL SPACING<br />

CHAN SPACING<br />

ADJ 20 kHz<br />

ALT1 40 kHz<br />

ALT2 60 kHz<br />

Note: Channel spacings can be selected separately for each channel. If a<br />

spacing value is changed in the table, the following values are<br />

changed automatically.<br />

IEC/IEEE-bus commands<br />

:[SENSe:]POWer:ACHannel:SPACing[:UPPer] 24KHZ<br />

:[SENSe:]POWer:ACHannel:SPACing:ACHannel 24KHZ<br />

:[SENSe:]POWer:ACHannel:SPACing:ALT 24KHZ<br />

EDIT<br />

ACP LIMITS<br />

The EDIT ACP LIMITS opens a table for defining the limits for ACPmeasurement.<br />

ACP LIMITS<br />

CHAN CHECK LOWER CHANNEL LIMIT UPPER CHANNEL LIMIT<br />

ADJ -50 dB -50 dB<br />

ALT1 -60 dB -60 dB<br />

ALT2<br />

The limit values are defined either in units of dB (forCP/ACP REL) or in dBm<br />

(for CP/ACP ABS).<br />

Note: Measured values that exceed one of the limits are marked with an<br />

preceding asterisk.<br />

IEC/IEEE-bus commands<br />

:CALCulate:LIMit:ACP:ACHannel 30DB, 30DB<br />

:CALCulate:LIMit:ACP:ACHannel:STATe ON|OFF<br />

:CALCulate:LIMit:ACP:ALT 30DB, 30DB<br />

:CALCulate:LIMit:ACP:ALT:STATe ON|OFF<br />

LIMIT<br />

CHECK<br />

The LIMIT CHECK softkey switches on or off the limit check of the ACp<br />

measurement.<br />

IEC/IEEE-bus commands<br />

:CALCulate:LIMit:ACPower[:STATe] ON | OFF<br />

:CALCulate:LIMit:ACP:ACHannel:RESult?<br />

:CALCulate:LIMit:ACP:ALTernate:RESult?<br />

% POWER<br />

BANDWIDTH<br />

The % POWER BANDWIDTH softkey opens an entry window for defining the<br />

percentage of the power with respect to the total power in the displayed<br />

frequency range. This defines the occupied bandwidth (percent of the total<br />

power).<br />

The permitted range is 10 % - 99,9 %<br />

IEC/IEEE-bus command :[SENSe:]POWer:BANDwidth 99PCT<br />