ESIB<br />

SENSe Subsystem<br />

:[SENSe:]MSUMmary:REFerence <br />

This command selects the reference modulation.<br />

Parameter: := 0.001PCT to 1000PCTfor AM<br />

0.1HZ to 10MHZfor FM<br />

0.0001RAD to 1000RADfor PM<br />

Example: ":MSUM:REF 50PCT"<br />

Features: *RST value: 100PCTfor AM<br />

100KHZfor FM<br />

10RAD for PM<br />

SCPI: device-specific<br />

Mode:<br />

VA-A<br />

:[SENSe:]MSUMmary:REFerence:AUTO ONCE<br />

This command sets the current absolute measured values of the main modulation signal as<br />

reference values for the relative indication.<br />

Example: ":MSUM:REF:AUTO ONCE"<br />

Features: *RST value: --<br />

SCPI: device-specific<br />

Mode:<br />

VA-A<br />

This command is an event and has therefore no query and no *RST value assigned.<br />

:[SENSe:]MSUMmary:MTIMe 0.1 s | 1 s<br />

This command selects the measurement time for the summary markers.<br />

Example: ":MSUM:MTIM 100US"<br />

Features: *RST value: 0.1S<br />

SCPI: device-specific<br />

Mode:<br />

VA-A<br />

1088.7531.12 6.141 E-15