14.06.2014 Views

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Macros<br />

ESIB<br />

DEFINE<br />

PAUSE<br />

The DEFINE PAUSE softkey inserts a pause instruction into the macro<br />

sequence which is being recorded. A running macro will pause when it<br />

reaches this instruction, at which point settings on the unit under test could for<br />

example be made. CONTINUE in the message window is used to continue<br />

the macro execution.<br />

IEC/IEEE-bus command --<br />

DELETE<br />

MACRO<br />

The DELETE MACRO softkey deletes the macro which was previously<br />

selected by the SELECT MACRO softkey. The softkey of the deleted title now<br />

carries its default label (MACRO X, X = macro number)<br />

IEC/IEEE-bus command --<br />

MACRO<br />

TITLE<br />

The MACRO TITLE softkey activates the entry of the title for the selected<br />

macro.<br />

Because the macro title is used for labelling the corresponding softkey, a<br />

maximum of only 20 characters is allowed. All text after the 10 th character is<br />

written to the second line. However, entering the character @ forces a new<br />

line at any point in the softkey title.<br />

IEC/IEEE-bus command --<br />

SELECT<br />

MACRO<br />

The SELECT MACRO softkey open the selection box with a list of all 7<br />

macros. All softkey functions of this menu are related to this macro.<br />

Macro 1<br />

(Macro 2)<br />

(Macro 3)<br />

(Macro 4)<br />

(Macro 5)<br />

(Macro 6)<br />

(Macro 7)<br />

SELECT MACRO<br />

IEC/IEEE-bus command --<br />

1088.7531.12 4.74 E-15