CALCulate Subsystem<br />

ESIB<br />

CALCulate:UNIT Subsystem<br />

The CALCulate:Unit subsystem defines the units for vector signal analyzer mode and power<br />

measurements.<br />

CALCulate<br />

:X<br />

:UNIT<br />

:TIME<br />

:UNIT<br />

:ANGLe<br />

:POWer<br />

COMMAND PARAMETERS UNIT COMMENT<br />

S | SYM<br />

DEG | RAD<br />

DBM | V | W | DB |<br />

PCT | UNITLESS |<br />

DBPW | WATT |<br />

DBUV | DBMV | VOLT |<br />

DBPT | DBUA | AMPere<br />

DBUV_MHZ | DBMV_MHZ |<br />

DBUA_MHZ | DBUV_M | DBUA_M |<br />

DBUV_MMHZ | DBUA_MMHZ<br />

Vector Signal Analysis<br />

Vector Signal Analysis<br />

:CALCulate:X:UNIT:TIME S | SYM<br />

This command selects seconds or symbols as an x-axis unit.<br />

Example: ":CALC:X:UNIT:TIME S"<br />

Features: *RST value: S<br />

SCPI: device-specific<br />

Mode:<br />

VA-D<br />

:CALCulate: UNIT:ANGLe DEG | RAD<br />

This command selects the unit for angular measurement.<br />

Example: ":CALC:UNIT:ANGL DEG"<br />

Features: *RST- value: RAD<br />

SCPI: device-specific<br />

Mode:<br />

VA-D<br />

:CALCulate: UNIT:POWer DBM | V | W | DB | PCT | DBPT | UNITLESS | DBPW | WATT |<br />

DBUV |DBMV | VOLT | DBUA | AMPere | DBUV_MHZ |DBMV_MHZ|<br />

DBUA_MHZ | DBUV_M | DBUA_M |DBUV_MMHZ | DBUA_MMHZ<br />

This command selects the unit for power.<br />

Example: ":CALC:UNIT:POW DBM"<br />

Features: *RST value: _<br />

SCPI: device-specific<br />

Modes:<br />

R, A, VA<br />

In the receiver mode, the units DBM, DBUV, DBPT, DBUA, DBPW, DBT, DBUV_M and DBUA_M<br />

are available. In the analyzer mode, all unit are available except for DBPT, DEG, RAD, S, and HZ.<br />

The units DEG, RAD, S, and HZ are available only in the vector analyzer mode.<br />

DBUV_MHZ and DBUA_MHZ denote the units DBUV/MHZ or DBUA/MHZ.<br />

1088.7531.12 6.58 E-15