SENSe Subsystem<br />

ESIB<br />

SENSe:DETector Subsystem<br />

The SENSe:DETector subsystem controls the recording of measurement values via the type of detector<br />

selected for each trace.<br />

The suffix in SENSe is not significant in this subsystem.<br />

[SENSe]<br />

:DETector<br />

[:FUNCtion]<br />

COMMAND PARAMETERS UNIT COMMENT<br />

:AUTO<br />

:RECeiver<br />

[:FUNCtion]<br />

:FMEasurement<br />

APEak | NEGative| POSitive | SAMPle| RMS<br />

| AVERage | QPEak | ACVideo<br />

<br />

POSitive | NEGative | RMS | AVERage |<br />

QPEak | ACVideo, POSitive | NEGative|<br />

RMS | AVERage | QPEak | ACVideo<br />

POSitive | NEGative | RMS | AVERage |<br />

QPEak | ACVideo<br />

--<br />

ACVideo only with option<br />

ESIB-B1<br />

:[SENSe:]DETector[:FUNCtion] APEak | NEGative | POSitive| SAMPle | RMS |<br />

AVERage | QPEak<br />

This command switches the detector for recording of the measured value.<br />

Example: ":DET POS"<br />

Features: *RST value: Trace1: POSitive<br />

Trace 2: AVERage<br />

SCPI: conforming<br />

Modes:<br />

R, A<br />

In scan mode of the receiver, the detectors POSitive, RMS, AVERage, QPEak and ACVideo are<br />

available (ACVideo only with option ESIB-B1).<br />

In the analyzer mode, the detectors APEak, POSitive, NEGative, RMS, SAMPLe and AVERage are<br />

available. The value "APEak" (AutoPeak) displays both the positive peak value and the negative<br />

peak value when noise is present. The positive peak value is displayed when one signal is present.<br />

The trace is selected by means of the numeric suffix after DETector.<br />

:[SENSe:]DETector[:FUNCtion]:AUTO ON | OFF<br />

This command either couples the detector to the current trace setting or turns coupling off.<br />

Example: ":DET:AUTO OFF"<br />

Features: *RST value: ON<br />

SCPI: conforming<br />

Mode:<br />

A<br />

The trace is defined by the numeric suffix at DETector.<br />

1088.7531.12 6.120 E-15