DISPlay Subsystem<br />

ESIB<br />

:DISPlay[:WINDow]:TRACe:Y[:SCALe]:RPOSition 0 to 100PCT<br />

This command defines the position of the reference value.<br />

Example: ":DISP:TRAC:Y:RPOS 50PCT"<br />

Features: *RST value: 100PCT (tracking generator)<br />

50PCT (vector analyzer)<br />

SCPI: conforming<br />

Modes:<br />

A, VA<br />

The numeric suffix in TRACe is not significant. This command is only valid in conjunction<br />

with option Tracking Generator or in vector analyzer mode.<br />

:DISPlay[:WINDow]:TRACe:Y[:SCALe]:PDIVision <br />

This command defines the scaling of the Y-axis in the current unit.<br />

Example: ":DISP:TRAC:Y:PDIV +1.20"<br />

Features: *RST value: -<br />

SCPI: conforming<br />

Mode:<br />

VA<br />

The numeric suffix in TRACe is not significant.<br />

:DISPlay[:WINDow]:TRACe:Y[:SCALe]:BOTTom <br />

This command defines the minimum grid level for the display in the receiver mode.<br />

Example: ":DISP:TRAC:Y:BOTT -20"<br />

Features: *RST value: 0<br />

SCPI: conforming<br />

Mode:<br />

R<br />

The numeric suffix in TRACe is not significant.<br />

:DISPlay[:WINDow]:TRACe:Y[:SCALe]:TOP <br />

This command defines the maximum grid level for the display in the receiver mode.<br />

Example: ":DISP:TRAC:Y:TOP 120"<br />

Features: *RST value: 100<br />

SCPI: conforming<br />

Mode:<br />

R<br />

The numeric suffix in TRACe is not significant.<br />

:DISPlay[:WINDow]:TRACe:Y:SPACing LINear | LOGarithmic | PERCent<br />

This command toggles between linear and logarithmic display.<br />

Example: ":DISP:TRAC:Y:SPAC LIN"<br />

Features: *RST value: LOGarithmic<br />

SCPI: conforming<br />

Mode:<br />

A<br />

The linear display can be LIN/% (PERCent) or LIN/dB (LINear). The numeric suffix in TRACe<br />

is not significant.<br />