ESIB<br />

INITiate Subsystem<br />

INITiate Subsystem<br />

The INITiate subsystem checks the initialization of the trigger subsystem.<br />

In receiver mode, a distinction is made between single measurement (INITiate1) and scan (INITiate2).<br />

In signal analyzer mode, a distinction is made between INITiate1 (screen A) and INITiate2 (screen B) in<br />

split screen representation.<br />

INITiate<br />

:CONTinuous<br />

:CONMeas<br />

[:IMMediate]<br />

:DISPlay<br />

COMMAND PARAMETERS UNIT COMMENT<br />

<br />

--<br />

--<br />

<br />

--<br />

--<br />

--<br />

--<br />

no query<br />

no query<br />

:INITiate:CONTinuous ON | OFF<br />

This command determines if the trigger system is continuously initiated ("Free Run").<br />

Example: ":INIT:CONT OFF"<br />

Features: *RST value: ON<br />

SCPI: conforming<br />

Modes:<br />

R, A, VA<br />

Setting "INITiate:CONTinuous ON" corresponds to function SCAN/SWEEP CONTinuous, ie the<br />

scan/sweep of the receiver/analyzer is cyclically repeated. The setting "INITiate:CONTinuous<br />

OFF" corresponds to function SCAN/SWEEP SINGLE.<br />

:INITiate:CONMeas<br />

This command continues the sweep from the current sweep position.<br />

Example: ":INIT:CONM"<br />

Features: *RST value: -<br />

SCPI: device-specific<br />

Modes:<br />

R, A, VA<br />

This command is an event and therefore has no *RST value and no query.<br />

Sweeps are stopped, for example, in the case of transducer sets between the different transducer<br />

factors.<br />

:INITiate[:IMMediate]<br />

The command initiates a new sweep/scan or starts a single sweep/scan.<br />

Example: "INIT"<br />

Features: *RST value: -<br />

SCPI: conforming<br />

Modes:<br />

R, A, VA<br />

This command is an event which is why it is not assigned an *RST value and has no query.<br />

:INITiate:DISPlay ON | OFF<br />

This command switches the display on or off during a single sweep.<br />

Example: ":INIT:DISP OFF"<br />

Features: *RST value: ON<br />

SCPI: device-specific<br />

Modes:<br />

R, A, VA<br />