Calibration<br />

ESIB<br />

<strong>EMI</strong><br />

PRESEL<br />

The <strong>EMI</strong> PRESEL softkey calibrates the frequency response of the<br />

preselector and the preamplifier.<br />

IEC/IEEE- bus command ::CALibration:PRESelector?<br />

PRESEL<br />

PEAK<br />

The PRESEL PEAK softkey optimizes the tuning of the preselector for input<br />

signals in the frequency range above 7 GHz.<br />

This calibration should be used whenever signal levels in the frequency range<br />

above 7 GHz are to be measured and high accuracy is desired.<br />

If a marker is not active when the PRESEL PEAK softkey is pressed, marker<br />

1 is activated as the reference marker and placed on the signal maximum in<br />

the active trace. Otherwise the active marker is used.<br />

The following window is displayed on the screen while the peaking function is<br />

running. The function can be aborted at any time using the ABORT line. In<br />

this case, the correction value determined at the factory is restored.<br />

SYSTEM MESSAGE<br />

PEAKING<br />

ABORT<br />

The input signal’s signal-to-noise ratio must be at least 10dB in order that the<br />

peaking function runs properly. Otherwise, the preselector may be set<br />

incorrectly, which would cause level errors by subsequent measurements.<br />

If the instrument setting (start/stop frequency, sweep time) is changed after<br />

PRESEL PEAK has been called, the correction value for the preselector<br />

determined by the function is no longer used. Instead, the correction value<br />

determined at the factory is restored.<br />

IEC/IEEE- bus command<br />

:CALibration:PPEak?<br />

CAL CORR<br />

ON OFF<br />

The CAL CORR ON/OFF softkey switches the correction data on/off.<br />

ON The status message depends upon the results of the total correction<br />

OFF The message UNCAL appears in the ESIB status line.<br />

IEC/IEEE- bus command :CALibration:STATe ON | OFF<br />

1088.7531.12 4.12 E-15