14.06.2014 Views

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

ESIB<br />

SENSe Subsystem<br />

:[SENSe:]SWEep:GAP:PRETrigger 0 to 100s<br />

This command defines the evaluation time for measured values before the pretrigger time<br />

(resolution: 50 ns).<br />

Example: ":SWE:GAP:PRET 100us"<br />

Features: *RST value: 0s<br />

SCPI: device-specific<br />

Mode:<br />

A<br />

:[SENSe:]SWEep:GAP:TRGTogap 0 to 100s<br />

This command defines the time between the pretrigger time and the beginning of the gap (trigger-togap<br />

time) (resolution: 50 ns).<br />

Example: ":SWE:GAP:TRGT 50us"<br />

Features: *RST value: 0s<br />

SCPI: device-specific<br />

Mode:<br />

A<br />

:[SENSe:]SWEep:GAP:LENGth 0 to 100s<br />

This command defines the gap length.<br />

Example: ":SWE:GAP:LENG 400us"<br />

Features: *RST value: 0s<br />

SCPI: device-specific<br />

Mode:<br />

A<br />

The gap length can be programmed from 0 to 100s in steps of 1, 2, 3 and 5.<br />

:[SENSe:]SWEep:SPACing LINear | LOGarithmic | AUTO<br />

This command toggles between linear and logarithmic step modes of the receiver. In analyzer mode,<br />

it toggles between linear and logarithmic sweep.<br />

Example: ":SWE:SPAC LOG"<br />

Features: *RST value: LIN<br />

SCPI: conforming<br />

Modes:<br />

R, A<br />

The frequency axis is set to linear or logarithmic scaling accordingly.<br />

1088.7531.12 6.153 E-15

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!