ESIB<br />

CALCulate Subsystem<br />

:CALCulate:DELTamarker:MODE ABSolute | RELative<br />

This command switches over between relative and absolute input of frequency of the delta marker.<br />

Example:<br />

":CALC:DELT:MODE ABS"<br />

Features: *RST value: REL<br />

SCPI: device-specific<br />

Modes:<br />

R, A, VA<br />

In the RELative mode, the frequency of the delta marker is programmed relative to the reference<br />

marker. In the ABSolute mode, the frequency is defined by the absolute values.<br />

:CALCulate:DELTamarker:AOFF<br />

This command switches off all active delta markers.<br />

Example:<br />

":CALC:DELT:AOFF"<br />

Features: *RST value: -<br />

SCPI: device-specific<br />

Modes:<br />

R, A, VA<br />

:CALCulate:DELTamarker:TRACe 1 to 4<br />

This command assigns the selected delta marker to the indicated measuring curve.<br />

Example: ":CALC:DELT3:TRAC 2"<br />

Features: *RST value: -<br />

SCPI: device-specific<br />

Modes:<br />

R, A, VA<br />

:CALCulate:DELTamarker:X 0 to MAX (frequency | sweep time | symbols)<br />

This command positions the selected delta marker to the indicated frequency (span > 0) or time<br />

(span = 0). The query always returns the absolute value of frequency or time.<br />

Example:<br />

":CALC:DELT:X 10.7MHz"<br />

Features: *RST value: -<br />

SCPI: device-specific<br />

Modes:<br />

R, A, VA<br />

The SYM unit is only valid in Vector Signal Analysis mode.<br />

:CALCulate:DELTamarker:X:RELative?<br />

This command queries the frequency (span > 0) or time (span = 0) of the selected delta marker<br />

relative to the reference marker.<br />

Example: ":CALC:DELT:X:REL?"<br />

Features: *RST value: -<br />

SCPI: device-specific<br />

Modes:<br />

R, A, VA<br />

1088.7531.12 6.9 E-15