ESIB<br />

Analyzer - Traces<br />

On completion of averaging, ie when the number of sweeps defined with SWEEP COUNT has been<br />

reached, a sliding average given by the following formula is obtained with CONTINUOUS SWEEP<br />

(N −1)<br />

⋅TRACEn - 1 + Meas Value , where Trace = new trace<br />

TRACEn<br />

=<br />

N<br />

Trace old = old trace<br />

N = SWEEP COUNT<br />

The display "Sweep 200 of 200" then remains constant until there is a new start.<br />

In SINGLE SWEEP mode, SWEEP START initiates n single sweeps. The sweeps are stopped as soon<br />

as the selected number of sweeps has been reached. The number of the current sweep and the total<br />

number of sweeps are shown in the display: "Sweep 3 of 200".<br />

SWEEP<br />

COUNT<br />

The SWEEP COUNT softkey activates the entry of the number of sweeps<br />

used for averaging.<br />

The range for SWEEP COUNT is 0 through 32767. For 0, the ESIB performs<br />

a sliding average over 10 sweeps. For 1, no averaging takes place.<br />

The default setting is 10 sweeps. The programming naturally influences the<br />

sweep duration. The number of sweeps used for averaging is the same for<br />

all 4 traces.<br />

Note : The setting of the SWEEP COUNT in the trace menu is equivalent<br />

to the setting in the sweep menu.<br />

IEC/IEEE-bus command :[SENSe:]SWEep:COUNt 10<br />

MAX HOLD<br />

The MAX HOLD softkey activates the max hold mode.<br />

In this display mode, the ESIB saves for each sweep the largest of the<br />

previously stored/currently measured values in the trace memory. The<br />

detector is set automatically to MAX PEAK. In this way, the maximum value<br />

of a signal can be determined over several sweeps.<br />

This is especially useful for measurement of modulated or pulsed signals.<br />

The signal spectrum is built up sweep by sweep until all the signal<br />

components have been captured.<br />

Pressing of the MAX HOLD softkey again clears the trace memory and<br />

restarts the max hold mode.<br />

If MAX HOLD is enabled, a new start is made after clearing the trace memory<br />

for each frequency change, (start frequency, stop frequency, center<br />

frequency or frequency span), a reference level change or switching between<br />

linear/logarithmic scales.<br />

IEC/IEEE-bus command<br />

:DISPlay[:WINDow]:TRACe:MODE MAXH<br />

1088.7531.12 4.229 E-15