SYSTem Subsystem<br />

ESIB<br />

:SYSTem:COMMunicate:GPIB[:SELF]:RTERminator LFEOI | EOI<br />

This command changes the delimiter.<br />

Example: ":SYST:COMM:GPIB:RTER EOI"<br />

Features: *RST value: LFEOI<br />

SCPI: device-specific<br />

Modes:<br />

R, A, VA<br />

The instrument contains a DMA-channel for communication via IEC-bus. This ensures maximum<br />

speed for the transfer of commands and data. The parser for command decoding integrated in the<br />

instrument is, however, only activated by the delimiter when the command is completely transferred.<br />

In order to make this possible for the transfer of binary data, too (e. g. trace data which are<br />

retransferred into the instrument), the delimiter recognition must be switched to the EOI signal prior<br />

to the transfer. Output of binary data from the instrument does not require such a switching.<br />

:SYSTem:COMMunicate:GPIB:RDEVice:ADDRess 0 to 30<br />

This command changes the IEC/IEEE-bus address of the unit which is selected as hardcopy device<br />

1 or 2, provided that the IEC/IEEE-bus interface of this unit is set as an interface.<br />

Example: ":SYST:COMM:GPIB:RDEV2:ADDR 5"<br />

Features: *RST value: 4<br />

SCPI: conforming<br />

Modes:<br />

R, A, VA<br />

:SYSTem:COMMunicate:SERial:CONTrol:DTR IBFull | OFF<br />

:SYSTem:COMMunicate:SERial:CONTrol:RTS IBFull | OFF<br />

These commands switch the hardware handshake procedure for the given serial interface off (OFF)<br />

or on (IBFull).<br />

Examples:<br />

":SYST:COMM:SER:CONT:DTR OFF"<br />

":SYST:COMM:SER2:CONT:RTS IBF"<br />

Features: *RST value: OFF<br />

SCPI: conforming<br />

Modes:<br />

R, A, VA<br />

The two commands have the same meaning. SERial1 and SERial 2 correspond to device COM1and<br />

COM2, respectively.<br />

:SYSTem:COMMunicate:SERial[:RECeive]:BAUD 75 | 150 | 300 | 600 | 1200 | 2400 | 9600<br />

This command sets the transmission speed for the given serial interface.<br />

Example: ":SYST:COMM:SER:BAUD 2400"<br />

Features: *RST value: 9600<br />

SCPI: conforming<br />

Modes:<br />

R, A, VA<br />

SERial1 and SERial 2 correspond to device interface COM1 and COM2, respectively. Permissible<br />

values are 75 Baud, 150 Baud, 300 Baud, 600 Baud, 1200 Baud, 2400 Baud, 4800 Baud, 9600<br />

Baud.<br />

1088.7531.12 6.168 E-15