Contents - Maintenance and Instrument Interfaces<br />

ESIB<br />

Handshake<br />

Software handshake<br />

In the software handshake mode of operation, the data transfer is controlled using the two control<br />

characters XON / XOFF.<br />

The instrument uses the control character XON to indicate that it is ready to receive data. If the receive<br />

buffer is full, it sends the XOFF character via the interface to the controller. The controller then interrupts<br />

the data output until it receives another XON from the instrument. The controller indicates to the<br />

instrument that it is ready to receive data in the same way.<br />

Cable required for software handshake<br />

The connection of the instrument with a controller for software handshake is made by crossing the data<br />

lines. The following wiring diagram applies to a controller with a 9-pin or 25-pin configuration.<br />

Instrument<br />

Controller<br />

DSR<br />

RxD<br />

RTS<br />

TxD<br />

CTS<br />

DTR<br />

GND<br />

1<br />

2<br />

3<br />

4<br />

5<br />

6<br />

7<br />

8<br />

9<br />

1<br />

2<br />

3<br />

4<br />

5<br />

6<br />

7<br />

8<br />

9<br />

DSR<br />

RxD<br />

RTS<br />

TxD<br />

CTS<br />

DTR<br />

GND<br />

DSUB connector female / 9 poles<br />

DSUB connector female / 9 poles<br />

Instrument<br />

Controller / PC<br />

DSR<br />

RxD<br />

RTS<br />

TxD<br />

CTS<br />

DTR<br />

GND<br />

1<br />

2<br />

3<br />

4<br />

5<br />

DSUB connector female / 9 poles<br />

6<br />

7<br />

8<br />

9<br />

1<br />

14<br />

2 TxD<br />

15<br />

3 RxD<br />

16<br />

4 RTS<br />

17<br />

5 CTS<br />

18<br />

6 DSR<br />

19<br />

7 GND<br />

20 DTR<br />

8<br />

21<br />

9<br />

22<br />

10<br />

23<br />

11<br />

24<br />

12<br />

25<br />

13<br />

DSUB connector female / 25 poles<br />

Fig. 8-3<br />

Wiring of the data lines for software handshake<br />

1088.7531.12 8.8 E-16