TRACe Subsystem<br />

ESIB<br />

TRACe Subsystem<br />

The TRACe subsystem controls access to the instrument’s internal trace memory.<br />

COMMAND PARAMETERS UNIT COMMENT<br />

TRACe<br />

[:DATA]<br />

:COPY<br />

:FEED<br />

:CONTrol<br />

TRACE1 | TRACE2 | TRACE3 | TRACE4 | SINGle |<br />

SCAN | STATus, |FINAL1|FINAL2|FINAL3|FINAL4,<br />

|...<br />

TRACE1|TRACE2|TRACE3|TRACE4,<br />

TRACE1|TRACE2|TRACE3|TRACE4<br />

ALWays | NEVer<br />

-<br />

:TRACe[:DATA] TRACE1| TRACE2| TRACE3| TRACE4| SINGle, | | FINAL1<br />

| FINAL2 | FINAL3 | FINAL4<br />

This command transfers trace data from the controller to the instrument, the query reads trace data<br />

out of the instrument.<br />

Example: ":TRAC TRACE1,"+A$ (A$: data list in the current format)<br />

"TRAC? TRACE1"<br />

Features: *RST value: -<br />

SCPI: conforming<br />

Modes:<br />

R, A, VA<br />

The trace data are transferred in the current format (corresponding to the command FORMat<br />

ASCii|REAL). The device-internal trace memory is addressed using the trace names ’TRACE1’ to<br />

’TRACE4’.<br />

The parameter of the query is the trace name TRACE1 to TRACE4, it indicates which trace memory<br />

will be read out.<br />

The transfer of trace data from the controller to the instrument takes place by indicating the trace<br />

name and then the data to be transferred. In ASCII format, these data are values separated by<br />

commas. If the transfer takes place using the format real (REAL 32), the data are transferred in block<br />

format (see FORMat subsystem).<br />

Saving and recalling trace data together with the device settings to/from the device-internal hard disk<br />

or to/from a floppy is controlled via the commands ":MMEMory:STORe:STATe" and<br />

":MMEMory:LOAD:STATe" respectively. Trace data are selected with<br />

":MMEMory:SELect[:ITEM]:ALL" or "":MMEMory:SELect[:ITEM]:TRACe". Trace data in<br />

ASCII format (ASCII FILE EXPORT) are exported with the command ":MMEM:STORe:TRACe".<br />

The transfer format for the trace data depends on the instrument setting<br />

Analyzer mode (span >0 and zero span):<br />

500 results are output in the unit selected for display.<br />

Note: With AUTO PEAK detector, only positive peak values can be read out.<br />

Trace data can be read into instrument with logarithmic display only in dBm, with<br />

linear display only in volts.<br />

FORMAT REAL,32 is to be used as format for binary transmission.<br />

1088.7531.12 6.174 E-15