ESIB<br />

Calibration<br />

Correction Function Calls<br />

SYSTEM CAL menu:<br />

CAL<br />

SHORT<br />

The CAL SHORT softkey starts a short calibration, in which the absolute gain<br />

of the analyzer as well as the gain errors of the selected bandwidth are<br />

corrected.<br />

IEC/IEEE- bus command :CALibration:SHORt?<br />

CAL<br />

TOTAL<br />

The CAL TOTAL softkey starts a complete correction of the receiver including<br />

additional partial corrections shown in the menu.<br />

UNCAL is displayed in the status line if the correction data recording has<br />

failed or if the correction values are deactivated (CAL CORR softkey= OFF).<br />

IEC/IEEE- bus command :CALibration[:ALL]?<br />

CAL<br />

RES BW<br />

The CAL RES BW softkey starts the correction of the center frequency, the<br />

bandwidth and the gain of the resolution filter.<br />

IEC/IEEE- bus command :CALibration:BANDwidth[:RESolution]?<br />

CAL LOG<br />

The CAL LOG softkey starts the correction of the linearity of the logarithmic<br />

amplifier.<br />

IEC/IEEE- bus command :CALibration:LDETector?<br />

CAL<br />

LO SUPP<br />

The CAL LO SUPP softkey starts the correction of the compensation of the<br />

first oscillator at low frequencies. After correction, the display of the internal<br />

oscillator at the frequency 0 Hz is very small.<br />

It is recommended that the correction be performed whenever sensitive<br />

measurements are to be made at low frequencies.<br />

CAL I/Q<br />

The CAL I/Q softkey starts the correction of the gain error of the I/Q<br />

demodulator.<br />

The softkey is only available in conjunction with option FSE-B7.<br />

IEC/IEEE- bus command :CALibration:IQ?<br />