ESIB<br />

Receiver - Frequency and Span<br />

Frequency and Span Selection – FREQUENCY Key Group<br />

The FREQUENCY key group is used to specify the receiver frequency and the frequency axis of the<br />

active measurement . The frequency axis is defined by the start and stop frequency.<br />

After pressing one of the keys FREQ, START or STOP, the value of the corresponding parameter can<br />

be defined in the appropriate dialog window.<br />

Start Frequency – START Key<br />

FREQUENCY – START menu:<br />

FREQUENCY<br />

CENTER/<br />

FREQ<br />

START<br />

SPAN/<br />

ZOOM<br />

STOP<br />

The START key opens a window where the start frequency of the scan<br />

diagram can be entered.<br />

The start frequency can be selected in the ESIB frequency range<br />

irrespective of scan range limits.<br />

IEC/IEEE bus command :[SENSe:]FREQuency:STARt <br />

Stop Frequency – STOP Key<br />

FREQUENCY – STOP menu:<br />

FREQUENCY<br />

CENTER/<br />

FREQ<br />

START<br />

SPAN/<br />

ZOOM<br />

STOP<br />

The STOP key opens a window where the stop frequency of the scan<br />

diagram can be entered.<br />

The stop frequency can be selected in the ESIB frequency range<br />

irrespective of scan range limits.<br />

The permissible value range for the stop frequency is:<br />

f stop > f start<br />

IEC/IEEE bus command :[SENSe:]FREQuency:STOP <br />

Receiver Frequency – FREQ Key<br />

FREQUENCY – FREQ menu:<br />

FREQUENCY<br />

CENTER/<br />

FREQ<br />

START<br />

SPAN/<br />

ZOOM<br />

STOP<br />

The FREQ key opens a window where the receiver frequency can be<br />

entered.<br />

A precondition is that the frequency and level readout is active.<br />

The permissible value range depends on the selected input:<br />

Input 1: 20 Hz ≤ f in ≤ f max<br />

Input 2: 20 Hz ≤ f in ≤ 1 GHz<br />

If the tuning frequency is lower than twice the IF bandwidth, the IF bandwidth<br />

is automatically reduced so that this condition is met again.<br />

If the frequency is increased again, the original IF bandwidth is restored. The<br />

memory is cleared when the IF bandwidth is manually changed.<br />

The resolution of the receive frequency is always 0.1 Hz.<br />

IEC/IEEE bus command :[SENSe:]FREQuency:STOP <br />