ESIB<br />

Calibration<br />

Results of System Error Correction<br />

SYSTEM CAL menu:<br />

CAL<br />

RESULTS<br />

The CAL RESULTS softkey in the right-hand supplementary menu calls the<br />

CALIBRATION RESULTS table, which displays the correction data that has<br />

been recorded.<br />

The CALIBRATION RESULTS table contains the following information:<br />

page 1:<br />

• date and time of the last recorded correction values<br />

• overall results of the correction value recording<br />

• list of the correction procedures according to function/module, including the<br />

correction values, the measured results and the individual results for each<br />

of the correction procedures<br />

The results for the individual correction procedures are categorised and<br />

displayed as follows:<br />

PASSED system error correction was completely successful<br />

CHECK deviations larger than expected, however, corrections could<br />

be performed<br />

FAILED deviations too large, no corrections were possible<br />

ABORTED correction aborted<br />

CALIBRATION RESULTS<br />

CALIBRATION: PASSED<br />

Last cal total: 05.Jun 1997 16:24:54<br />

Calibration of IF Filters PASSED<br />

IF GAIN Adjust<br />

PASSED<br />

Bandwidth:<br />

Filter Cal Val [Hz] DAC Val State<br />

1kHz 2.806e+01 1679 PASSED<br />

2kHz 1.603e+01 2887 PASSED<br />

3kHz -6.012e+00 3238 PASSED<br />

5kHz -1.002e+01 3514 PASSED<br />

10kHz 1.804e+02 3703 PASSED<br />

20kHz 3.607e+02 3801 PASSED<br />

30kHz 8.417e+02 3831 PASSED<br />

50kHz 1.403e+03 3743 PASSED<br />

100kHz 1.804e+03 3698 PASSED<br />

200kHz 3.607e+03 3606 PASSED<br />

300kHz 8.417e+03 3516 PASSED<br />

500kHz 1.403e+04 3329 PASSED<br />

1MHz 2.806e+04 2881 PASSED<br />

IEC/IEEE- bus command --<br />

PAGE UP<br />

The PAGE UP and PAGE DOWN softkeys can be used to page through the<br />

table of correction results.<br />

PAGE DOWN<br />

1088.7531.12 4.13 E-15