14.06.2014 Views

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

ESIB<br />

Instrument Interfaces<br />

User Interface (USER)<br />

The user interface, located on the rear panel of the ESIB, is a 25 pin Cannon connector which provides<br />

access to the two user ports (Port A and Port B). Each port is 8 bits wide (A0 - A7 and B0 -B7) and can<br />

be configured either as output or as input. The voltage levels are TTL levels (Low < 0,4 V, High > 2 V).<br />

In addition, an internal +5 V power supply voltage is provided. The maximum load current is 100 mA.<br />

The pin assignments for the USER connector can be seen in the following diagram:<br />

GND GND A7<br />

+5 V GND GND A6<br />

A5<br />

A4<br />

A3<br />

A2<br />

A1<br />

A0<br />

13<br />

25<br />

1<br />

14<br />

+5 V<br />

B7<br />

B6<br />

B5<br />

B4<br />

B3<br />

B2 B0<br />

B1<br />

Fig. 8-5<br />

Pin assignments for the USER connector.<br />

The configuration of the user ports takes place in the SETUP menu (SETUP key) in the GENERAL<br />

SETUP sub-menu.<br />

1088.7531.12 8.21 E-16