ESIB<br />

Analyzer - Traces<br />

The TRACE keys 1...4 call a menu whith the setting options for the selected trace.<br />

This menu is used to determine the method for compressing the measurement data in the frequency or<br />

time domain to the 500 representable points of the display is determined.<br />

When the measurement is started, a new trace can be displayed or the trace can be based on previous<br />

results. The traces can be displayed, blanked and copied.<br />

Traces can also be corrected with the mathematical functions.<br />

The measurement detector for each of the display types may be chosen either directly or selected<br />

automatically by the ESIB.<br />

All activated traces are marked with a LED at the corresponding key (here, TRACE 1). The default<br />

setting is TRACE 1 with CLEAR / WRITE selected. The remaining traces 2...4 are switched off<br />

(BLANK). For the split screen display, the selection of the trace automatically selects the corresponding<br />

screen for entry.<br />

Only one of the CLEAR/WRITE, MAX HOLD, MIN HOLD, AVERAGE, VIEW and BLANK softkeys can<br />

be activated at any one time.<br />

CLEAR/<br />

WRITE<br />

The CLEAR/WRITE softkey activates the clear/write display mode.<br />

The trace is displayed without additional trace evaluation. The trace memory<br />

is overwritten by each sweep. If more than one data point falls within a pixel,<br />

the trace is displayed in bar form with the maximum and minimum values<br />

being assigned to one pixel. In the clear/write display mode, all the available<br />

detectors are selectable. The autopeak detector is selected in the default<br />

mode (detector to AUTO).<br />

Whenever the CLEAR/WRITE softkey is pressed, the ESIB clears the<br />

selected trace memory and restarts the measurement.<br />

IEC/IEEE-bus command<br />

:DISPlay[:WINDow]:TRACe:MODE WRITe<br />

VIEW<br />

The VIEW softkey freezes the current contents of the trace memory and<br />

displays them.<br />

If the trace data were formed through MAX HOLD, MIN HOLD or AVERAGE,<br />

the sweep is restarted and the trace contents are cleared, after switching to<br />

these trace modes.<br />

If a trace is frozen by VIEW, the instrument settings can be modified without<br />

modifying the displayed trace. The fact that the trace and the current<br />

instrument setting do not agree anymore is indicated by an enhancement<br />

label "*" at the right edge of the grid. The initial instrument setting can be<br />

restored using the ADJUST TO TRACE softkey in the TRACE MATH submenu.<br />

If LEVEL RANGE or REF LEVEL is changed in the VIEW display mode, the<br />

ESIB adjusts the measurement data to the changed display range. Thus, an<br />

amplitude zoom can be performed after the measurement so that details of<br />

the trace can be seen better.<br />

IEC/IEEE-bus command<br />

:DISPlay[:WINDow]:TRACe:MODE VIEW<br />

1088.7531.12 4.227 E-15