I<br />

Start-Up<br />

ESIB<br />

AC power switch on the rear panel<br />

ON<br />

OFF<br />

0<br />

Ac power switch<br />

Fuse holder<br />

AC power connector<br />

Power switch<br />

After turn-on (position ON), the instrument is in the ready<br />

state (STANDBY) or in operation dependent upon the<br />

position of the ON/STANDBY switch located on the<br />

instrument front panel (see below).<br />

Note: The AC power switch can be left ON all the time..<br />

Switching to OFF is only required when the instrument<br />

has to be completely disconnected from the AC power<br />

source.<br />

Switching to OFF (position OFF) disconnects the ESIB<br />

completely from the AC power source.<br />

ON/STANDBY switch on the front panel<br />

ON<br />

STANDBY<br />

.<br />

Standby<br />

½ ON/STANDBY switch is not depressed.<br />

The yellow LED (STANDBY) is on. Only the ESIB‘s<br />

internal power is connected to the AC line. The crystal<br />

oven is maintained at the normal operating temperature.<br />

Operation<br />

Caution:<br />

In standby mode, the ESIB is<br />

still live.<br />

½ ON/STANDBY is depressed.<br />

The green LED (ON) is on. The instrument is ready for<br />

operation. All modules within the instrument are supplied<br />

with power.<br />

Switching on the Instrument<br />

½ To switch on the ESIB, set the power switch on the rear panel to position I.<br />

½ Pressing the ON/STANDBY key on the front panel. The green LED ahould come on.<br />

Startup Menu and Booting<br />

When the instrument is switched on, a message indicating the installed BIOS version (eg Analyzer BIOS<br />

Rev. 1.2) appears on the screen for a few seconds.<br />

Windows NT is booted first and then the instrument firmware. As soon as boot up is over, the instrument<br />

will start measuring. The settings used will be the one that was active when the instrument was<br />

previously switched off, provided no other device configuration than FACTORY had been selected with<br />

AUTO RECALL in the MEMORY menu.<br />

Switching off the Instrument<br />

½ Remove the disk, if any, from the drive before switching off.<br />

½ Press the ON/STANDBY key on the front panel. The yellow LED should come on.<br />

Only when completely disconnecting the ESIB from the AC power source:<br />

½ Set the power switch on the rear panel to position 0.<br />

1088.7531.12 1.20 E-3