Example - Level and Frequency Measurement<br />

ESIB<br />

DEFINE<br />

SCAN<br />

3. Programming a SCAN table<br />

½ Press the DEFINE SCAN key.<br />

A menu is displayed where the whole display range<br />

can be defined and divided into scan subranges.<br />

After PRESET a scan range of 150 kHz to 1 GHz is<br />

set on the frequency axis. The SCAN table is<br />

automatically activated.<br />

Now the stop frequency is set to 30 MHz.<br />

DEFINE<br />

SCAN<br />

USER<br />

SCAN<br />

SCAN TABLE<br />

Start<br />

Stop<br />

Step<br />

150 kHz<br />

1 GHz<br />

LIN Auto<br />

Max Level<br />

Min Level<br />

Frequency Axis LOG<br />

100 dBµV<br />

0 dBµV<br />

ADJUST<br />

AXIS<br />

SINGLE<br />

SCAN<br />

CONTINUOUS<br />

SCAN<br />

RANGE 1<br />

Start 150.000 kHz<br />

Stop<br />

30 MHz<br />

Step Size 4 kHz<br />

RES BW 9 kHz<br />

Meas Time 1 ms<br />

Auto Ranging OFF<br />

RF Attn 10 dB<br />

Preamp OFF<br />

Auto Preamp OFF<br />

Input INPUT 1<br />

RANGE 2 RANGE 3 RANGE 4 RANGE 5<br />

30 MHz<br />

1 GHz<br />

40 kHz<br />

120 kHz<br />

100 µs<br />

OFF<br />

10 dB<br />

OFF<br />

OFF<br />

INPUT 1<br />

SCAN RANGES<br />

SCAN<br />

RANGES<br />

INS BEFORE<br />

RANGE<br />

INS AFTER<br />

RANGE<br />

DELETE<br />

RANGE<br />

RANGES<br />

1-5 6-10<br />

RUN<br />

SCAN<br />

Fig.2-2 Scan range after selecting default setup with PRESET<br />

:<br />

DATA VARIATION<br />

½ Press the DOWN cursor key in the DATA<br />

VARIATION field.<br />

The set stop frequency is marked.<br />

1088.7531.12 2.4 E-3