14.06.2014 Views

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

ESIB<br />

Receiver - Main Markers<br />

Marker Step Size<br />

DATA VARIATION – STEP menu:<br />

DATA VARIATION<br />

HOLD STEP<br />

MARKER<br />

STEP SIZE<br />

STEPSIZE<br />

AUTO<br />

STEPSIZE<br />

MANUAL<br />

The STEP key in the DATA VARIATION key group opens a<br />

menu to match the marker step size to each individual<br />

application. In order to change the step size, the marker entry<br />

mode must already be active.<br />

Control is returned to the MARKER NORMAL menu via the<br />

menu key .<br />

MKR TO<br />

STEPSIZE<br />

DELTA TO<br />

STEPSIZE<br />

.<br />

.<br />

.<br />

STEPSIZE<br />

AUTO<br />

STEPSIZE<br />

MANUAL<br />

MKR TO<br />

STEPSIZE<br />

DELTA TO<br />

STEPSIZE<br />

The STEPSIZE AUTO softkey sets the marker step size to AUTO. In this<br />

case, the step size is exactly 10% of the grid. A one-raster movement of the<br />

roll-key corresponds to one pixel.<br />

STEPSIZE AUTO is the default setting.<br />

IEC/IEEE bus command<br />

:CALCulate:MARKer:STEP:AUTO ON|OFF<br />

The STEPSIZE MANUAL softkey activates the entry window for defining a<br />

fixed value for the marker step size.<br />

Pressing the step key shifts the marker position by the selected step size.<br />

The resolution of the roll-key is always one pixel per raster.<br />

IEC/IEEE bus command<br />

:CALCulate:MARKer:STEP 10KHZ<br />

The MKR TO STEPSIZE softkey sets the marker step size to the current<br />

marker frequency.<br />

This function is well suited to harmonic measurements. The marker is set to<br />

the largest signal using the peak search function. After activation of the MKR<br />

TO STEPSIZE function, the marker is set to the corresponding harmonic of<br />

the signal each time the cursor key or is pressed when entering the<br />

marker position.<br />

IEC/IEEE bus command<br />

:CALCulate:MARKer:FUNCtion:MSTep<br />

The DELTA TO STEPSIZE softkey sets the marker step size to the<br />

difference between the reference marker and the last active delta marker.<br />

The softkey is only available when at least one delta marker is enabled.<br />

IEC/IEEE bus command --<br />

1088.7531.12 4.123 E-15

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!