Sweep Control - Analyzer<br />

ESIB<br />

SWEEPTIME<br />

AUTO<br />

SWEEPTIME<br />

MANUAL<br />

The SWEEPTIME AUTO and SWEEPTIME MANUAL softkeys activate the<br />

automatic or manual selection of the sweep time. These functions are<br />

identical to the entries in the COUPLING menu (see section "Setting and<br />

Coupling the Coupling Resolution, Video Bandwidth and Sweep Time").<br />

IEC/IEEE-bus commands<br />

:[SENSe:]SWEep:TIME:AUTO ON | OFF<br />

:[SENSe:]SWEep:TIME 10s<br />

SGL SWEEP<br />

DISP OFF<br />

The SGL SWEEP DISP OFF softkey switches off the display during a single<br />

sweep. The trace is displayed when the sweep has been completed.<br />

IEC/IEEE-bus commands<br />

:INITiate:DISPlay ON | OFF;<br />

:INITiate<br />

SWEEP<br />

COUNT<br />

The SWEEP COUNT softkey activates the entry of the number of sweeps<br />

performed by the ESIB after the start of a single sweep. If trace average, max<br />

hold or min hold is switched on, this also determines the number of averages<br />

or minimua/maxima that are taken.<br />

Example:<br />

[TRACE1: MAX HOLD]<br />

[SWEEP: SWEEP COUNT: {10} ENTER]<br />

[SINGLE SWEEP]<br />

The ESIB performs the max hold function over 10 sweeps.<br />

The sweep count range is 0 through 32767. For sweep count = 0 or 1, a<br />

sweep is performed. For trace averaging (AVERAGE), sweep count = 0 and<br />

continuous sweep, the ESIB performs a sliding average over 10 sweeps in<br />

average mode. For sweep count = 1, no averaging takes place.<br />

The default setting totals 10 sweeps.<br />

Note: The setting for the number of sweeps in the TRACE menu is<br />

equivalent to the setting in the SWEEP menu. In SINGLE SWEEP<br />

mode, the measurement is stopped after reaching the chosen<br />

number of sweeps.<br />

IEC/IEEE-bus command :[SENSe:]SWEep:COUNt 0<br />