14.06.2014 Views

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

ESIB<br />

Analyzer - Main Markers<br />

0<br />

-10<br />

-20<br />

-30<br />

Ref Lvl<br />

-20 dBm<br />

Marker [T1]<br />

-28.4 dBm<br />

200.0100 MHz<br />

C0<br />

RBW 300 Hz<br />

VBW 300 Hz<br />

SWT 100 ms<br />

C0<br />

1<br />

RF ATT 10 dB<br />

Mixer -30 dBm<br />

Unit dBm<br />

1 -28.4 dBm<br />

200.0100 MHz<br />

CH0 PWR -18.3 dBm<br />

ACP Max -67.23 dBm<br />

ACP UP -67.23 dBm<br />

ACP LOW -69.45 dBm<br />

CH Space 30 kHz<br />

Ch BW<br />

24 kHz<br />

-40<br />

-50<br />

-60<br />

-70<br />

-80<br />

-90<br />

-100<br />

Center<br />

200 MHz<br />

10 kHz/<br />

Span 100 kHz<br />

Channel BW<br />

Channel BW<br />

Channel BW<br />

Channel<br />

Spacing<br />

Channel<br />

Spacing<br />

Fig.4-11<br />

Lower<br />

Channel<br />

Transmit<br />

Channel<br />

Measurement of the adjacent channel power.<br />

Upper<br />

Channel<br />

If the limit check is activated for ACP measurements, the result of the limit check (PASSED/FAILED) is<br />

displayed as well as an asterisk preceding each measured ACP power that exceeded one of the limits.<br />

Note:<br />

The limit check is activated and the limits are defined in the POWER MEAS SETTINGS menu.<br />

1088.7531.12 4.193 E-15

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!