CALCulate Subsystem<br />

ESIB<br />

CALCulate:PEAKsearch - Subsystem<br />

The CALCulate:PEAKsearch subsystem allows data to be processed for final measurement.<br />

CALCulate<br />

Command PARAMETER Unit Comment<br />

:PEAKsearch | :PSEarch<br />

[:IMMediate]<br />

:MARGin<br />

:SUBRanges<br />

:METHod<br />

<br />

<br />

SUBRange | PEAK<br />

dB<br />

--<br />

:CALCulate:PEAKsearch|PSEarch[:IMMediate]<br />

This command activates the generation of final measurement data.<br />

Example: "CALC:PEAK"<br />

Features: *RST value: -<br />

SCPI: device-specific<br />

Mode:<br />

R<br />

The numeric suffix in CALCULATE is not significant.<br />

:CALCulate:PEAKsearch|PSEarch:MARGin MINimun .. MAXimum<br />

This command defines the margin of the final measurement.<br />

Example: "CALC:PEAK:MARG 5 dB"<br />

Features: *RST value: 6 dB<br />

SCPI: device-specific<br />

Mode:<br />

R<br />

The numeric suffix in CALCULATE is not significant.<br />

:CALCulate:PEAKsearch|PSEarch:SUBRanges 1 to 500<br />

This command defines the number of subranges for the final measurement.<br />

Example: "CALC:PEAK:SUBR 10"<br />

Features: *RST value: 25<br />

SCPI: device-specific<br />

Mode:<br />

R<br />

The numeric suffix in CALCULATE is not significant.<br />

1088.7531.12 6.56 E-15