ESIB<br />

Analyzer - Marker Å Menu<br />

Instrument Parameter Changes via Markers – MKR Í Key<br />

MARKER MKR →menu:<br />

NORMAL<br />

DELTA<br />

MARKER<br />

SEARCH<br />

MKR<br />

MARKER-><br />

PEAK<br />

MKR-><br />

CENTER<br />

MKR-><br />

REF LEVEL<br />

The MKR → menu offers functions that can be used to<br />

change instrument settings with the currently active marker.<br />

Exactly as in the SEARCH menu, these functions can also be<br />

applied to the delta markers.<br />

The choice between marker and delta marker is made<br />

according to the currently active frequency entry mode for the<br />

marker/delta marker. If no entry mode is active, the marker<br />

with the lowest number will be activated as the reference<br />

marker.<br />

MKR->CF<br />

STEPSIZE<br />

MKT-><br />

START<br />

MKR-><br />

STOP<br />

MKR-><br />

TRACE<br />

SELECT<br />

MARKER<br />

ACTIVE<br />

MKR DELTA<br />

ACTIVE<br />

MKR DELTA<br />

The ACTIVE MKR /DELTA softkey toggles between the active marker and<br />

the active delta marker.<br />

If DELTA is illuminated, the following marker functions are carried out with the<br />

active delta marker.<br />

Note: Switching between marker and delta marker can also be done with<br />

the NORMAL and DELTA keys.<br />

IEC/IEEE-bus command --<br />

PEAK<br />

To simplify control, the PEAK search function (see section "Search Functions<br />

- SEARCH Key") is also available in the MRK→ menu. Thus, the most<br />

important functions MARKER→PEAK, MKR→CENTER and MKR→REF<br />

LEVEL can be selected in one menu.<br />

IEC/IEEE-bus commands<br />

:CALCulate:MARKer:MAX<br />

:CALCulate:DELTamarker:MAX<br />

MKR-><br />

CENTER<br />

The MKR→CENTER softkey sets the center frequency to the current<br />

marker/delta marker frequency .<br />

The softkey is not available in the time domain.<br />

IEC/IEEE-bus command<br />

:CALCulate:MARKer:FUNCtion:CENTer<br />

1088.7531.12 4.211 E-15