ESIB<br />

SENSe Subsystem<br />

:[SENSe:]FILTer:LPASs:FREQuency <br />

This command defines the frequency limit of the low-pass filter in the NF-branch for analog<br />

demodulation.<br />

Parameter: ::= 3 kHz | 15 kHz for REAL TIME ON<br />

5 PCT | 10 PCT | 25 PCT for REAL TIME OFF<br />

Example: ":FILT:LPAS:FREQ 3KHZ" for REAL TIME ON<br />

":FILT:LPAS:FREQ 25PCT" for REAL TIME OFF<br />

Features: *RST value: - (STATe = OFF)<br />

SCPI: conforming<br />

Mode:<br />

VA-A<br />

:[SENSe:]FILTer:CCITt[:STATe] ON | OFF<br />

This command activates the weighting filter according to CCITT-recommendation in the AF-branch<br />

for analog demodulation.<br />

Example: ":FILT:CCIT ON"<br />

Features: *RST value: OFF<br />

SCPI: conforming<br />

Mode:<br />

VA-A<br />

This command is valid only in vector analysis mode with REAL TIME OFF.<br />

:[SENSe:]FILTer:CMESsage[:STATe] ON | OFF<br />

This command activates the C-message weighting filter according to CCITT-recommendation in the<br />

AF-branch for analog demodulation.<br />

Example: ":FILT:CMES ON"<br />

Features: *RST value: OFF<br />

SCPI: conforming<br />

Mode:<br />

VA-A<br />

This command is valid only in vector analysis mode with REAL TIME OFF.<br />

:[SENSe:]FILTer:DEMPhasis[:STATe] ON | OFF<br />

This command activates the selected de-emphasis for analog demodulation.<br />

Example: ":FILT:DEMP ON"<br />

Features: *RST value: OFF<br />

SCPI: conforming<br />

Mode:<br />

VA-A<br />

1088.7531.12 6.131 E-15