ESIB<br />

Analyzer - Display and Limit Lines<br />

DISPLAY<br />

LINE 1<br />

DISPLAY<br />

LINE 2<br />

The DISPLAY LINE 1/2 softkeys switch the display lines on/off and activate<br />

the entry of the line location.<br />

The display lines mark the selected levels in the measurement window.<br />

IEC/IEEE-bus commands<br />

:CALCulate:DLINe:STATe ON | OFF;<br />

:CALCulate:DLINe -20dBm<br />

THRESHOLD<br />

LINE<br />

The THRESHOLD LINE softkey switches the threshold line on/off and<br />

activates the entry of the line location.<br />

The threshold line is a display line which defines a threshold value. This<br />

threshold value serves as a lower search limit for maximums/minimums in<br />

the marker functions (MAX PEAK, MIN PEAK, NEXT PEAK etc.). For signal<br />

tracking (SIGNAL TRACK function), this threshold value defines the lower<br />

search limit (see section "Marker Functions")..<br />

IEC/IEEE-bus commands<br />

:CALCulate:THReshold ON | OFF;<br />

:CALCulate:THReshold -82dBm<br />

REFERENCE<br />

LINE<br />

The REFERENCE LINE softkey switches the reference line on/off and<br />

activates the entry of the line position.<br />

The reference line serves as the basis for mathematical correlation between<br />

trace data (see section "Mathematical Operations on Measurement Curves"<br />

IEC/IEEE-bus commands<br />

:CALCulate:RLINe:STATe ON | OFF;<br />

:CALCulate:RLINe -10dBm<br />

FREQUENCY<br />

LINE 1<br />

FREQUENCY<br />

LINE 2<br />

The FREQUENCY LINE 1/2 softkeys switch the frequency lines 1/2 on/off<br />

and activate the entry of the line locations.<br />

The frequency lines mark the selected frequencies in the measurement<br />

window or define search ranges (see section "Marker Functions").<br />

IEC/IEEE-bus commands<br />

:CALCulate:FLINe:STATe ON | OFF;<br />

:CALCulate:FLINe 120 MHz<br />