Traces - Receiver<br />

ESIB<br />

TRACE 1 menu<br />

TRACE<br />

1 2<br />

TRACE 1<br />

CLEAR/<br />

WRITE<br />

VIEW<br />

TRACE 1<br />

T1-T2+REF<br />

->T1<br />

T1-T3+REF<br />

->T1<br />

3<br />

3<br />

4<br />

BLANK<br />

T1-T4+REF<br />

->T1<br />

T1-REF<br />

->T1<br />

MAX HOLD<br />

MIN HOLD<br />

ASCII<br />

EXPORT<br />

Traces 3 and 4 only<br />

FINAL<br />

RESULTS<br />

SCAN<br />

COUNT<br />

ASCII<br />

CONFIG<br />

DETECTOR<br />

COPY<br />

TRACE MATH<br />

OFF<br />

The TRACE keys 1...4 call a menu which presents the setting options for the selected trace.<br />

This menu is used to determine how the measured data will be represented in the frequency domain<br />

using the 500 pixels provided by the display.<br />

When the measurement is started, a new trace can be displayed or the trace can be based on previous<br />

results.<br />

Traces can be displayed, blanked and copied. By applying mathematical functions, the traces can also<br />

be corrected. The measurement detector for each of the display types may be chosen directly by the<br />

ESIB.<br />

All activated traces are marked with a LED at the corresponding key (here, TRACE 1). The default<br />

setting is TRACE 1 with CLEAR / WRITE and the PEAK detector selected, TRACE2 with CLEAR /<br />

WRITE and the AVERAGE detector selected . The remaining traces 3 to 4 are switched off (BLANK).<br />

The CLEAR/WRITE, MAX HOLD, MIN HOLD, VIEW, and BLANK are mutually exclusive selection<br />

switches.<br />

CLEAR/<br />

WRITE<br />

The CLEAR/WRITE softkey activates the clear/write display mode.<br />

The trace is displayed without additional trace evaluation. The trace memory<br />

is overwritten by each scan. If more than one data point falls within a pixel,<br />

the trace is displayed in bar form with the maximum and minimum values in a<br />

pixel connected. In the clear/write display mode, all the available detectors<br />

are selectable.<br />

Whenever the CLEAR/WRITE softkey is pressed, the ESIB clears the<br />

selected trace memory and restarts the measurement.<br />

IEC\IEEE bus command<br />

:DISPlay[:WINDow]:TRACe:MODE WRITe<br />