YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

ESIB- Receiver<br />

ESIB<br />

Structure of the ASCII file:<br />

<strong>RECEIVER</strong> mode, final measurement data:<br />

Content of file<br />

File header Type;ESIB 7;<br />

Version;2.07;<br />

Date;01.Jan 2000;<br />

Mode;Receiver;<br />

Start;10000;Hz<br />

Stop;100000;Hz<br />

x-Axis;LIN;<br />

Scan Count;1;<br />

Transducer;TRD1;<br />

Scan 1:<br />

Start;150000;Hz<br />

Stop;1000000;Hz<br />

Step;4000;Hz<br />

RBW;100000;Hz<br />

Meas Time;0.01;s<br />

Auto Ranging;ON;<br />

RF Att;20;dB<br />

Auto Preamp;OFF;<br />

Preamp;0;dB<br />

Input;1;<br />

Description<br />

Instrument model<br />

Firmware version<br />

Date record storage date<br />

Instrument operating mode<br />

Start/stop of the display range.<br />

Unit: Hz<br />

Scaling of x axis linear (LIN) or logarithmic (LOG)<br />

Number of scans set<br />

Transducer name (if switched on)<br />

Loop over all defined scan ranges (1-10)<br />

Range - start frequency in Hz<br />

Range – stop frequency in Hz<br />

Range - step width in Hz for linear step width or in %<br />

(1-100) for logarithmic step width<br />

Range - resolution bandwidth<br />

Range - measurement time<br />

Auto ranging on - or off for current range<br />

Range - input attenuation<br />

Auto Preamp on or off for current range<br />

Range - preamplifier on (20dB) or off (0dB)<br />

Range - input (1 or 2)<br />

Data section<br />

of the file<br />

TRACE 1 FINAL:<br />

Trace Mode;AVERAGE;<br />

Final Detector<br />

x-Unit;Hz;<br />

y-Unit;dBuV;<br />

Final Meas<br />

Time;1.000000;s<br />

Margin;6.000000:s<br />

Values;8;<br />

2;154000.000000;81.638<br />

535;15.638535;N;GND<br />

1;158000.000000;86.563<br />

789;7.563789;N;GND<br />

2;1018000.000000;58.68<br />

9873;-1.310127;N;GND<br />

...<br />

Selected trace<br />

Trace mode:<br />

CLR/WRITE,AVERAGE,MAX HOLD,MIN HOLD,<br />

VIEW, BLANK<br />

Final detector<br />

MAX PEAK, MIN PEAK, RMS, AVERAGE,<br />

QUASI PEAK, AC VIDEO<br />

Unit of x values:<br />

Unit of y values:<br />

Final measurement time<br />

Margin<br />

Number of test points<br />

Measured values:<br />

;, ; ; <br />

Phase and protective grounding are output only if a<br />

line impedance stabilization has been activated. They<br />

specify the setting at which the maximum RFI level at<br />

the associated frequency was found.<br />

1088.7531.12 4.102 E-15