14.06.2014 Views

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

el. level<br />

Pulse frequency/Hz<br />

Fig. 3<br />

Differences of the weighting of pulsed sinusoidal signals resulting from display modes AV,<br />

CISPR AV and Pk depending on the pulse width (measurement time = 2 s,<br />

pulse width = 10 ms, IF bandwidth = 9 kHz, averaging time constant = 160 ms).<br />

rel. level/dB<br />

Pulse frequency/Hz<br />

Fig. 4<br />

Differences of the weighting of pulsed sinusoidal signals resulting from display modes AV,<br />

CISPR AV and Pk depending on the pulse frequency (measurement time = 2 s,<br />

pulse width = 10 ms, IF bandwidth = 120 kHz, averaging time constant = 100 ms).<br />

1088.7490.01 E E-10