Contents - Maintenance and Instrument Interfaces<br />

ESIB<br />

RS-232-C Interface<br />

The standard ESIB is equipped with two serial interfaces (RS-232-C).The interfaces can be set up<br />

manually in the SETUP-GENERAL SETUP menu in the COM PORT1/2 table.<br />

Each of the active RS-232-C interfaces is assigned to one of the 9-pin connectors located on the rear<br />

panel. Interface 1 is assigned to the connector COM1 and Interface 2 is assigned to the connector<br />

COM2.<br />

Interface Characteristics<br />

• Serial data transmission in asynchronous mode<br />

• Bidirectional data transfer via two separate lines<br />

• Transmission rate selectable from 110 to 19200 baud<br />

• Logic ’0’ signal from+3 V to +15 V<br />

• Logic ’1’ signal from -15 V to -3 V<br />

• An external instrument (controller) can be connected.<br />

• Software handshake (XON, XOFF)<br />

• Hardware handshake<br />

RxD DTR<br />

DCD TxD SG<br />

1 2 3 4 5<br />

6 7 8 9<br />

RTS RI<br />

DSR CTS<br />

Fig. 8-2<br />

Pin assignment of the RS-232-C interface<br />

Signal Lines<br />

1. Data lines<br />

The data transmission is bit-serial in ASCII code starting with the LSB. Two lines, RxD and TxD are<br />

necessary as the minimum requirement for transmission; however, no hardware handshake is<br />

possible. For handshaking, only the XON/XOFF software handshake protocol can be used .<br />

RxD (Receive Data)<br />

Input, LOW = logic ’1’, HIGH = logic ’0’.<br />

Data line, local terminal receives data from remote station.<br />

TxD (Transmit Data)<br />

Output, LOW = logic ’1’, HIGH = logic ’0’.<br />

Data line, local terminal transmits data to remote station.<br />

2. Control lines<br />

DCD (Data Carrier Detect),<br />

Not used in ESIB.<br />

DTR (Data Terminal Ready),<br />

Output, active LOW,<br />

Indicates that the local terminal is ready to receive data.<br />

1088.7531.12 8.6 E-16