Coupled Settings - Analyzer<br />

ESIB<br />

RBW / VBW<br />

PULSE [.1]<br />

The RBW / VBW PULSE softkey sets the following coupling ratio:<br />

video bandwidth = 10 x resolution bandwidth<br />

or<br />

video bandwidth = 10 MHz (= maximum video bandwidth).<br />

This coupling ratio is recommended when the amplitude of pulsed signals is<br />

to be measured. The IF filter characteristics alone determine the pulse shape.<br />

No additional weighting is provided by the video filter.<br />

This setting is only effective for the VBW AUTO selection in the main menu.<br />

IEC/IEEE-bus command :[SENSe:]BWIDth:VIDeo:RATio PULSe<br />

RBW / VBW<br />

NOISE [10]<br />

The RBW / VBW NOISE softkey sets the following coupling ratio:<br />

video bandwidth = resolution bandwidth/10<br />

At this coupling ratio, noise and pulsed signals are suppressed in the video<br />

section. For noise signals, the ESIB displays the average value.<br />

This softkey is disabled unless VBW AUTO is selected in the main menu.<br />

IEC/IEEE-bus command<br />

:[SENSe:]BWIDth:VIDeo:RATio NOISe<br />

RBW / VBW<br />

MANUAL<br />

The RBW / VBW MANUAL softkey activates the entry of the coupling ratio for<br />

resolution bandwidth to video bandwidth.<br />

The range for the RBW/VBW ratio is 0,001 to 1000.<br />

This softkey is disabled unless VBW AUTO is selected in the main menu.<br />

IEC/IEEE-bus command :[SENSe:]BWIDth:VIDeo:RATio 10<br />

SPAN / RBW<br />

AUTO [50]<br />

The SPAN / RBW AUTO [50] softkey sets the following coupling (rounding is<br />

made to the next higher value):<br />

resolution bandwidth = span/50<br />

This coupling corresponds to the default state.<br />

This softkey is disabled unless RBW AUTO is selected in the main menu.<br />

IEC/IEEE-bus command<br />

:[SENSe:]BWIDth[:RESolution]:RATio 0.02<br />

SPAN / RBW<br />

MANUAL<br />

The SPAN / RBW MANUAL softkey activates the entry for the coupling of<br />

resolution bandwidth and span.<br />

The range for the SPAN/RBW ratio is 1 to 10000.<br />

This softkey is disabled unless RBW AUTO is selected in the main menu.<br />

IEC/IEEE-bus command<br />

:[SENSe:]BWIDth[:RESolution]:RATio 0.1<br />