Switchover to Remote Control<br />

ESIB<br />

Remote Control via RS-232-Interface<br />

Setting the Transmission Parameters<br />

To enable an error-free and correct data transmission, the parameters of the unit and the controller<br />

should have the same setting. Parameters can be manually changed in menu SETUP-GENERAL<br />

SETUP in table COM PORT 1/2 or via remote control using the command<br />

SYSTem:COMMunicate:SERial1|2:... .<br />

The transmission parameters of the interfaces COM1 and COM2 are factory-set to the following values:<br />

baudrate = 9600, data bits = 8, stop bits = 1, parity = NONE and owner = INSTRUMENT.<br />

<strong>Manual</strong>ly:<br />

Setting interface COM1|2<br />

½ Call SETUP-GENERAL SETUP menu<br />

½ Select desired baudrate, bits, stopbit, parity in table COM PORT 1/2.<br />

½ Terminate input using one of the unit keys (=ENTER).<br />

Return to <strong>Manual</strong> Operation<br />

Return to manual operation is possible via the front panel or via RS-232 interface.<br />

<strong>Manual</strong>ly:<br />

½ Press the LOCAL key.<br />

Notes: Before switchover, command processing must be completed as<br />

otherwise switchover to remote control is effected immediately.<br />

– The LOCAL key can be disabled by the universal command LLO<br />

(see chapter 8) in order to prevent unintentional switchover. In this<br />

case, switchover to manual mode is only possible via remote<br />

control.<br />

– The LOCAL key can be enabled again by sending the control<br />

codes "@LOC" via RS-232 (see chapter 8).<br />

Via RS-232: ...<br />

V24puts(port, ’@LOC’); Set instrument to manual operation.<br />

...<br />

Limitations<br />

The following limitations apply if the unit is remote-controlled via the RS-232-C interface:<br />

− No interface messages, some control codes are defined (see chapter 8).<br />

− Only the Common Commands *OPC? can be used for command synchronization, *WAI and *OPC<br />

are not available.<br />

− Block data cannot be transmitted.<br />

When Windows NT is booted, data are output via the COM interface because of automatic external<br />

device recognition. Therefore, it is recommended to clear the input buffer of the controller before remote<br />

operation of the instrument via the COM interface.<br />

1088.7531.12 5.4 E-16