14.06.2014 Views

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

ESIB- Receiver<br />

ESIB<br />

Selection of Detectors for Final Measurement<br />

The selection of detectors for the final measurement is made in the TRACE menu (see section<br />

"Selection and setting of traces ").<br />

The detectors to be used for the final measurement can be set here for each trace, ie any combination<br />

of prescan and final measurement is possible. The required flexibility is thus obtained for the diverse<br />

test specifications which are covered by means of the ESIB.<br />

Traces 3 and 4 can be used for displaying the final measurement values. For this purpose, the FINAL<br />

MEAS softkey should be activated in the associated menu. Trace 3 represents the allocated final<br />

measurement values referring to trace 1 with the symbol "x" and trace 4 the final measurement values<br />

for trace 2 with "+".<br />

The results of prescan and final measurement in a diagram:<br />

Att 0 dB AUTO<br />

INPUT 1<br />

100<br />

Det<br />

AV Trd<br />

ResBW 120 kHz<br />

Meas T 100 ms Unit<br />

100 MHz<br />

dBpT<br />

90<br />

80<br />

70<br />

1MA<br />

2AV<br />

60<br />

xQP<br />

+AV<br />

50<br />

40<br />

30<br />

014QP<br />

014AV<br />

20<br />

10<br />

0<br />

30 MHz 300 MHz<br />

Date: 14.DEC.1999 11:41:33<br />

Fig. 4-8<br />

Results of prescan and final measurement<br />

1088.7531.12 4.108 E-15