14.06.2014 Views

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

ESIB<br />

Analyzer - Search Functions<br />

MARKER SEARCH menu<br />

SUMMARY<br />

MARKER<br />

SUMMARY<br />

MARKER<br />

The SUMMARY MARKER softkey calls up the submenu for<br />

selecting the summary marker measurements.<br />

The softkey is only available in the time domain.<br />

RMS<br />

MEAN<br />

PEAK HOLD<br />

ON OFF<br />

AVERAGE<br />

ON OFF<br />

SWEEP<br />

COUNT<br />

ALL SUM<br />

MKR OFF<br />

RMS<br />

MEAN<br />

The RMS softkey selects the measurement of the rms value of the signal per<br />

sweep.<br />

In case of maximum peak evaluation the highest rms value since the activation of<br />

PEAK HOLD ON is displayed. In case of AVERAGE ON the rms values of a trace<br />

are averaged and displayed over several sweeps.<br />

The number of sweeps is set with the SWEEP COUNT softkey. If PEAK HOLD =<br />

ON is active at the same time, the display is held until the occurrence of the next<br />

higher average value.<br />

IEC/IEEE-bus commands<br />

:CALCulate:MARKer:FUNCtion:SUMM:RMS ON<br />

:CALCulate:MARKer:FUNCtion:SUMM:RMS:RES?<br />

:CALCulate:MARKer:FUNCtion:SUMM:RMS:AVER:RES?<br />

:CALCulate:MARKer:FUNCtion:SUMM:RMS:PHOL:RES?<br />

The MEAN softkey selects the measurement of the average value of the signal per<br />

sweep.<br />

Thus, the mean power can be measured, for example during a GSM burst.<br />

In case of maximum peak evaluation, the highest average value since the<br />

activation of PEAK HOLD ON is displayed.<br />

In case of AVERAGE ON, the average values of a trace are averaged and<br />

displayed over several sweeps.<br />

The number of sweeps is set with the SWEEP COUNT softkey. If PEAK HOLD =<br />

ON is active at the same time, the display is held until the occurrence of the next<br />

higher average value.<br />

IEC/IEEE-bus commands :<br />

:CALCulate:MARKer:FUNCtion:SUMM:MEAN ON<br />

:CALCulate:MARKer:FUNCtion:SUMM:MEAN:RES?<br />

:CALCulate:MARKer:FUNCtion:SUMM:MEAN:AVER:RES?<br />

:CALCulate:MARKer:FUNCtion:SUMM:MEAN:PHOL:RES?<br />

1088.7531.12 4.209 E-15