Contents - Maintenance and Instrument Interfaces<br />

ESIB<br />

Instrument Interfaces<br />

IEC Bus Interface<br />

The standard instrument is equipped with an IEC/IEEE Bus connector. An IEEE 488 interface connector<br />

is located on the rear panel of the ESIB. An external controller for remote control of the instrument can<br />

be connected via the IEEE 488 interface connector using a shielded cable.<br />

A further optional IEC/IEEE-bus interface can be installed in the ESIB and is assigned to the controller.<br />

This interface can be controlled by standard software (R&S-Basic, QUICK-Basic, etc.). It enables the<br />

remote control of the instrument via an external link to the two IEC/IEEE-bus interfaces and in addition<br />

the control of other equipment via the IEC/IEEE-bus connector at the rear of the unit (eg control of a<br />

complete test setup).<br />

The following section describes the first IEC Bus interface over which the instrument can be remotely<br />

controlled. The characteristics of the PC2A/PC-AT controller interface depend upon the user-installed<br />

software running on the controller and is, therefore, not described here.<br />

Interface Characteristics<br />

• 8-bit parallel data transfer<br />

• bidirectional data transfer<br />

• three-line handshake<br />

• high data transfer rate of max. 350 kbyte/s<br />

• up to 15 instruments can be connected<br />

• maximal length of the interconnecting cables 15 m (single connection, 2m)<br />

• wired-OR connection if several instruments are connected in parallel.<br />

ATN IFC NRFD EOI DIO3 DIO1<br />

SHIELD SRQ NDAC DAV DIO4 DIO2<br />

12 1<br />

13<br />

24<br />

LOGIC GND GND(10) GND(8) GND(6) DIO8 DIO6<br />

GND(11) GND(9) GND(7) REN DIO7 DIO5<br />

Fig. 8-1<br />

Pin assignment of IEC-Bus interface<br />

1088.7531.12 8.2 E-16