Sweep Control - Analyzer<br />

ESIB<br />

Ref Lev<br />

-10.0 dBm<br />

RBW<br />

VBW<br />

SWT<br />

100 kHz<br />

100 kHz<br />

500 us<br />

RF Att 20 dB<br />

Unit<br />

(dBm)<br />

Span 0 Hz 50 us / Div Center 914 MHz<br />

Fig. 4-19<br />

Display of a burst with a gap<br />

The GAP SWEEP measurement is activated by the GAP SWEEP ON/OFF softkey. The mode settings<br />

are made in the GAP SWEEP SETTINGS sub-menu.<br />

SWEEP SWEEP menu:<br />

GAP SWEEP<br />

ON OFF<br />

The GAP SWEEP ON/OFF softkey switches the GAP SWEEP measurement<br />

on/off.<br />

The softkey is available only in the time domain.<br />

IEC/IEEE-bus command :[SENSe:]SWEep:GAP ON | OFF<br />