ESIB Maintenance<br />

8 Maintenance and Instrument Interfaces<br />

The following chapter contains information on the maintenance of the ESIB and on the instrument<br />

interfaces.<br />

The address of our support center and a list of all <strong>Rohde</strong> & <strong>Schwarz</strong> service centers can be found at the<br />

beginning of this manual.<br />

Maintenance<br />

Mechanical Maintenance<br />

The ESIB does not require any mechanical maintenance. The front panel can be cleaned from time to<br />

time using a damp, soft cloth.<br />

Electrical Maintenance<br />

Testing the Level Measuring Accuracy<br />

Due to the total calibration facility by means of the built-in calibration generator, high long-term stability<br />

of the level measuring characteristics is guaranteed. It is recommended to check the measuring<br />

accuracy every two years according to the performance test. Reprogramming of the correction data by a<br />

R&S servicing shop is necessary, if tolerances are exceeded.<br />

Testing the Frequency Accuracy<br />

The frequency accuracy of the reference oscillator is to be checked once a year according to section 5.<br />

It is not necessary to perform this test if the instrument is operated using an external reference.<br />

Test utility:<br />

Test setup:<br />

Frequency counter of signal generator<br />

Measurement with frequency counter<br />

Connect frequency counter to the REF OUT/IN socket at the rear of the<br />

instrument.<br />

Measurement with signal generator<br />

Apply signal with 1GHz, -10dBm to RFin.<br />

Measurement: Settings on ESIB: MODE ANALYZER<br />

[CENTER 1000 MHz]<br />

[SPAN 0 MHz]<br />

[REF REF LEVEL -10 dBm]<br />

[MARKER COUNT ON<br />

COUNTER RESOL 0.1 Hz]<br />

Activate internal frequency counter (marker function)<br />

1088.7531.12 8.1 E-16