ESIB<br />

Configuration/Setup<br />

An overwritten transducer set remains stored in the background as long as the edited factor is stored<br />

with the SAVE TRD SET softkey or until the table is closed. A set overwritten by mistake can be<br />

restored by leaving the entry.<br />

TRANSD SET<br />

NAME<br />

The TRD SET NAME softkey activates the entry of the transducer set<br />

characteristics in the header field of the table.<br />

Name - Entry of name<br />

A maximum of 8 characters is permissible for the name. The characters have<br />

to comply with the convention of DOS file names. The unit automatically<br />

stores all transducer sets with the extension .TDS.<br />

If an existing name is changed, the set stored under the previous name is<br />

retained and will not be overwritten automatically with the new name. The<br />

previous set can be deleted at a later time using DELETE FACTOR/SET.<br />

Thus, sets can be copied.<br />

IEC/IEEE-bus command :[SENSe:]CORR:TSET:SELect <br />

TRANSD SET<br />

UNIT<br />

Unit - Selection of unit<br />

The unit of the transducer set is selected from a selection box activated by<br />

the TRANSD SET UNIT softkey.<br />

The unit should be selected prior to the entry as it determines the settable<br />

transducer factors. The preset unit for new sets is "dB". The unit can no<br />

longer be changed when editing a set as the set of selected transducer<br />

factors will otherwise become inconsistent.<br />

SET UNIT<br />

dB<br />

dBm<br />

dBµV<br />

dBµV/m<br />

dBµA<br />

dBµA/m<br />

dBpW<br />

dBpT<br />

IEC/IEEE-bus command<br />

:[SENSe:]CORR:TSET:UNIT <br />

Break - Activation of query when changing a subrange<br />

The sweep can be stopped if the user changes the subrange and selects a<br />

new subrange of the transducer. A message informs the user that the limit<br />

has been attained. He can continue the sweep or switch off the transducer.<br />

The interruption is activated by setting Break to ON. Selection is by the<br />

ENTER key which toggles between ON and OFF (toggle function).<br />

IEC/IEEE-bus command :[SENSe:]CORR:TSET:BREak ON|OFF<br />

Comment - Entry of a comment<br />

The comment can be freely selected. It can have a maximum number of 50<br />

characters.<br />

IEC/IEEE-bus command<br />

:[SENSe:]CORR:TSET:COMMent <br />

1088.7531.12 4.29 E-15